Low RNA yield or poor A260/230: causes and fixes

A poor A260/230 usually means salt or phenol carry-over rather than bad RNA. Low yield usually starts at lysis. Here is how to read the numbers and what to change.

Range: Nucleic acid extraction and purification3 minute readUpdated

Racks of capped sample tubes ready for extraction
On this page
  1. Key points
  2. What the numbers mean
  3. Fixing a low A260/230
  4. Fixing low yield
  5. Degraded RNA
  6. Genomic DNA carry-over
  7. Questions

Key points

  • A low A260/230 usually means guanidine salts, phenol or carbohydrate carried over. The RNA itself is often fine.
  • A low A260/280 points to protein or phenol. The ratio also depends on pH, so measure in a buffered solution.
  • Low yield usually starts at lysis: incomplete homogenisation or too much starting material.
  • Absorbance cannot show integrity. Check degradation on a gel or with an RNA integrity number.

What the numbers mean

Reading RNA absorbance ratios
ReadingExpected for pure RNAIf it is low
A260/280About 2.0Protein or phenol carry-over, or an acidic blank
A260/230About 2.0–2.2Guanidine salts, phenol, carbohydrate or ethanol carry-over

Ratios also become unreliable at low concentrations, roughly below 20 ng/µL, where small absorbance errors have a large effect. Use a fluorometric assay for dilute samples.

Fixing a low A260/230

  • Silica columns. Add one more wash with the ethanol-containing wash buffer, and spin the empty column for a minute to remove every trace of ethanol before eluting.
  • Keep the membrane clean. Do not let lysate touch the column rim, where it escapes the washes.
  • Phenol-based extractions. Avoid the interphase, and wash the pellet twice in 75% ethanol.
  • Plant and polysaccharide-rich samples. Use a kit or additive designed for them. See extraction by sample type.
  • Rescue. An ethanol or lithium chloride precipitation can clean up RNA that has already been eluted.

Fixing low yield

  • Incomplete lysis. Tough tissue needs a rotor-stator or bead homogeniser. Pieces left after lysis take their RNA with them.
  • Too much starting material. Overloading a column or the lysis buffer reduces yield instead of raising it. Stay within the kit’s stated input.
  • Binding conditions. Check that ethanol was added to the wash buffer on first use and to the lysate where the protocol calls for it.
  • Elution. Apply water or buffer directly to the centre of the membrane and leave it for a minute before spinning. A second elution with the first eluate can raise the concentration.

Chemistry differs between methods: the extraction methods guide compares column, bead and phenol approaches.

Degraded RNA

Intact total RNA from mammalian cells shows sharp 28S and 18S bands at about a 2:1 ratio on a gel. Smearing toward low molecular weights means degradation. A capillary instrument gives an RNA integrity number from 1 (degraded) to 10 (intact).

  • Freeze tissue immediately in liquid nitrogen or place it in a stabilisation reagent.
  • Keep frozen samples frozen until they are in lysis buffer.
  • Use RNase-free tips and tubes, wear gloves and clean the bench with an RNase remover.

Genomic DNA carry-over

A signal in the no-RT control of a qPCR run means genomic DNA came through. Include an on-column or in-solution DNase step and design primers across an exon junction. Kits with and without DNase steps are in the extraction kit range.

Questions we get asked

Why is my A260/230 low?

Usually carry-over of guanidine salts from the lysis or binding buffer, phenol, carbohydrate or ethanol. An extra wash and a dry spin before elution often fix column preps.

Is RNA with a low A260/230 still usable?

Often yes for robust assays, but salts and phenol can inhibit reverse transcription. Test it, or clean it up by precipitation if results are poor.

What is a good RNA integrity number?

The scale runs from 1 to 10, with 10 fully intact. The value you need depends on the application; RNA sequencing is generally more demanding than a short-amplicon RT-qPCR.

Why does my A260/280 change between measurements?

The ratio depends on pH and ionic strength. Measuring in water gives lower, more variable values than measuring in a buffer at about pH 7.5.

Sources

  1. Single-step method of RNA isolation by acid guanidinium thiocyanate-phenol-chloroform extraction. Chomczynski P et al. Anal Biochem. 1987;162:156-9.
  2. Effect of pH and ionic strength on the spectrophotometric assessment of nucleic acid purity. Wilfinger WW et al. Biotechniques. 1997;22:474-6, 478-81.
  3. The RIN: an RNA integrity number for assigning integrity values to RNA measurements. Schroeder A et al. BMC Mol Biol. 2006;7:3.

For research use only. Protocols here are general outlines: follow your kit manual and your institution’s safety rules. See our editorial policy.

Products for this workflow

Nucleic acid extraction and purification

DNA, RNA and plasmid kits, with sample-specific kits for blood, tissue, plant, fecal and viral RNA samples.

Need the reagents for this?

Send the product names or catalogue numbers, quantities and delivery country. We'll confirm availability, lead time and the documents for nucleic acid extraction and purification.