ELISA high background or no signal: causes and fixes

High background usually means washing or detection reagent; no signal usually means a step, a reagent or the sample. Here is how to tell which, using the wells you already ran.

Range: ELISA kits4 minute readUpdated

Pipette tip dispensing into a red microplate on a white bench
On this page
  1. Key points
  2. Read the plate before changing anything
  3. High background
  4. No signal or weak signal
  5. Poor duplicates or a poor standard curve
  6. When the kit may not be the right one
  7. Questions

Key points

  • High background across the whole plate usually means too little washing, too much detection reagent or a contaminated substrate.
  • No signal in any well, including the top standard, means a missed step or a dead reagent. Signal in the standards but not the samples means the samples.
  • Poor duplicates are nearly always pipetting, bubbles or uneven washing.
  • Very high samples can read falsely low in a sandwich ELISA. Dilute them and see whether the result rises.

Read the plate before changing anything

Your standards and blanks already say which half of the assay failed. Look at them before repeating the run.

What the plate is telling you
What you seeWhat it points toCheck first
Every well strongly coloured, blanks includedDetection or substrate problemWashing, conjugate dilution, substrate colour before use
No colour anywhere, top standard includedA step or reagent failedStep order, conjugate added, substrate activity
Standards fine, samples blankSample below range, or wrong sample typeKit range, dilution, validated matrices
Standards fine, samples above the top standardSample too concentratedDilute and rerun
Duplicates disagreeHandlingPipetting, bubbles, washer
Outer wells differ from inner wellsEdge effectTemperature, evaporation, plate sealing

High background

  • Too little washing. Residual conjugate is the commonest cause. Fill wells completely, soak briefly if the manual allows, and tap the plate dry on clean paper after the last wash.
  • A blocked washer pin. One clogged channel gives one streak of high wells. Test the washer with water and check every channel dispenses.
  • Too much detection antibody or conjugate. Recheck the dilution. Do not reuse a working solution from an earlier run.
  • Substrate problems. TMB should be almost colourless before use. If it is already blue, it has been contaminated or exposed to light.
  • Incubating too long or too warm. Keep to the stated time and temperature, and read soon after adding stop solution.
  • Carry-over. Use fresh plate sealers at each step and a fresh tip for each reagent. A trace of HRP on a tip is enough to colour a substrate reservoir.

No signal or weak signal

  • A step was missed or done out of order. Work through the manual with a tick sheet on the repeat run.
  • Sodium azide in a buffer. Azide inhibits HRP, so azide-preserved wash or sample buffers can abolish the signal.
  • Reagents not at room temperature. Cold reagents slow binding and colour development. Bring them up for about 30 minutes first.
  • Degraded standard or conjugate. Reconstitute as directed, aliquot, and avoid repeated freeze-thaw. Check expiry and storage.
  • Wrong wavelength. TMB with an acid stop reads at 450 nm. A reference at 570 or 630 nm is optional and corrects for plate imperfections.
  • Wrong species or sample type. A kit only measures what it was validated for. See how the formats differ.

Poor duplicates or a poor standard curve

  • Duplicates. Pre-wet tips, use the same technique for every well and remove bubbles before reading. Mix each sample before pipetting.
  • Standard curve. Mix every step of the dilution series thoroughly and use a fresh tip at each step, since an error at the top carries through all the points.
  • Curve fit. Immunoassay curves are sigmoidal. A four-parameter logistic fit usually describes them better than a straight line.
  • Edge effects. Bring the plate and reagents to temperature first, seal it well and avoid stacking plates in the incubator.

A sandwich ELISA can also show a high-dose hook: at extreme concentrations the target saturates both antibodies and the reading falls. If a sample reads lower than expected, rerun it at a higher dilution and see whether the calculated concentration goes up.

When the kit may not be the right one

If controls pass and your samples still fail, the kit may not suit the sample type or species. Send us the target, species and sample and we can check validated options in the ELISA kit range.

Questions we get asked

Why is my whole ELISA plate blue?

If every well turns blue quickly, blanks included, the usual causes are too little washing, too much conjugate or contaminated substrate. Check that the TMB was close to colourless before use and repeat with a fresh tip for every reagent.

Can sodium azide affect an ELISA?

Yes. Sodium azide inhibits horseradish peroxidase, so azide in buffers used near the HRP step can reduce or abolish the signal. Use azide-free buffers for HRP-based detection.

What causes poor duplicates in ELISA?

Most often inconsistent pipetting, bubbles in the wells, samples that were not mixed, or a washer with a partly blocked channel.

What is the hook effect in ELISA?

At very high analyte concentrations, a sandwich ELISA can give a falsely low reading because excess target saturates both capture and detection antibodies. Diluting the sample further reveals it.

Sources

  1. A Practical Guide to Immunoassay Method Validation. Andreasson U et al. Front Neurol. 2015;6:179.
  2. Heterophilic antibody interference in immunometric assays. Bolstad N et al. Best Pract Res Clin Endocrinol Metab. 2013;27:647-61.
  3. Enzyme-linked immunosorbent assay (ELISA). Quantitative assay of immunoglobulin G. Engvall E et al. Immunochemistry. 1971;8:871-4.

For research use only. Protocols here are general outlines: follow your kit manual and your institution’s safety rules. See our editorial policy.